Physics for the Life Sciences
Experiment 49
The Slide-wire Potentiometer
Purpose
To study the operation of a potentiometer circuit and to use it
to determine the emf and internal resistance of a dry cell.
Apparatus
Slide wire potentiometer Galvanometer Resistance box (0 - 100 W)
Power supply (6 to 12 V DC) Ammeter(1 A DC) Standard cell
Single pole, double throw switch
1½ V dry cell (D or C size)
Variable resistor (100 W) Connecting wires
Theory
Whenever a voltmeter is used to measure the potential difference
across part of an electrical circuit, the effective resistance of that
part of the circuit is altered and the current in it changes since some
of the current is needed to actuate the meter. These changes in the
circuit can be minimized by using a very high resistance voltmeter.
There are some instances, however, when we need to measure a potential
difference without drawing any current from the circuit at all. One of
these is the measurement of the electromotive force of a cell. This is the
potential difference across the terminals of the cell when no current is
taken from it; that is an open circuit. A potentiometer is a device
which allows measurements of potential difference without drawing any
current from the circuit. It is a null method. The circuit of a
potentiometer is shown below.
A simple potentiometer circuit consists of a uniform slide wire,
AB, usually 1 m long through which a constant current is maintained by
a power supply. The cell or battery whose emf is to be measured, Ex
is connected through a galvanometer, G, to the slider, C. A balance
condition exists when the potential difference across AC is just equal
to the emf of the cell, Ex in which case there will be no deflection on
the galvanometer. This procedure is then repeated with a standard cell
of known emf, C, so that the wire may be calibrated.
Suppose Ls is the length AC when the standard cell of
emf Es is used, and Lx is the length for the unknown
cell of emf, Ex, then Ex/Es =Lx/ Ls, or
Ex = Es(Lx/Ls)............................................................................................... (1)
The value of an unknown emf may then be determined from the measurements
of Lx and Ls, if the standard emf, Es is known.
It is often convenient to adjust the current through the wire AB
so that the potential difference across it is a certain value such as
exactly 1 V. In this case, each cm along the wire represents 0.01 V or
10 mV and each mm represents 1 mV. Commercial potentiometers are designed
this way so that the potential difference to be determined may be
measured directly on a dial calibrated in volts or millivolts.
When current is drawn from a cell or battery, the
potential difference across its terminals decreases as the current from it
increases. This is due to the internal resistance within the battery. The
difference between the emf of the cell on open circuit and the potential
difference across its terminals when delivering a current I, is then equal to
the potential difference across its internal resistance, Ir. Thus E- V = Ir, or
r =
(E - V)/I. The value of the internal resistance
of a cell may then be determined by measuring its emf, E,
and its terminal potential difference, V, while a current I is being drawn
from it. The current drawn from a cell of emf E when an
external resistance R is connected across it is given by Ohm's
law; I =
E/(R + r) where r is the internal resistance of the cell.
Hence r = (E - V)/I = (E - V)(R + r)/E, or, solving for R,
R = (R/V)E - r ........................................................................ (2)
This is a linear equation of R with R/V. A graph of R (on the y-axis) versus R/V (on the x-axis) should
then be linear with a slope equal to the emf, E. The negative intercept
on the R axis is equal to the internal resistance, r.
Procedure
- Connect up the circuit as shown with the key K open and the power
supply switched off. Have your instructor check your circuit.
- Switch on the power supply with the standard cell, Es, connected to the
circuit. Adjust the slider, C, to the 50 cm mark and adjust the
current through the potentiometer wire until there is a null reading
on the galvanometer when the key is depressed. Record the reading
of the ammeter current. This reading should be checked periodically
throughout the experiment and the variable resistor adjusted as needed
to keep this value constant. Record the emf of the standard cell in
the table.
- Close the switch K so as to connect the dry cell as the unknown, Ex, in
the circuit instead of the standard cell. Readjust the position of
the slider until a null reading on the galvanometer is found. Record
this position of the slider in the table.
- Leave the dry cell in the circuit as you have it now and connect the
resistance box in parallel with it. Adjust the value on the box
to 20W and determine the slider position for a null reading on the
galvanometer. Record the length AC.
- Decrease the value on the resistance box to 18W then in steps of 2W
down to 10W, and in steps of lW down to 3W .
In each case,
measure the slider position for a null reading on the galvanometer
and enter the data in the table.
- Disconnect the resistance box and remeasure Ex for the dry cell.
Calculations
- Determine the potential difference per cm along the potentiometer
wire and enter the result in the table.
- Calculate the emf of the dry cell from the value of Lx obtained in
procedure (3).
- Calculate the terminal voltage of the dry cell for each value of R in
procedure (5) and the corresponding values of R/V.
- Plot a graph of R versus R/V. Determine the emf of
the dry cell from the slope of the graph and the value of the internal
resistance from the negative intercept.
Data
Resistance R (W) |
Length Lx (cm) |
Terminal Voltage V
|
R/V (A-1) |
| 20 |
|
|
|
| 18 |
|
|
|
| 16 |
|
|
|
| 14 |
|
|
|
| 12 |
|
|
|
| 10 |
|
|
|
| 9 |
|
|
|
| 8 |
|
|
|
| 7 |
|
|
|
| 6 |
|
|
|
| 5 |
|
|
|
| 4 |
|
|
|
| 3 |
|
|
|
| infinite |
|
|
|
Current through galvanometer wire................................................................ ______ A
Emf of standard cell......................................................................................... ______ V
Balance length for dry cell (Lx)....................................................................... ______ cm
Emf of dry cell.................................................................................................. ______ V
Potential difference per cm along potentiometer........................................... ______ mV/cm
Emf of dry cell determined from slope of graph............................................. ______ V
Internal resistance of dry cell determined from graph.................................. ______ W
Emf of dry cell at end of experiment.............................................................. ______ V
Questions
- Compare the measured values of the emf of the dry cell obtained
procedure (3) with that determined from the slope of the graph.
Explain any differences in the values. How does the value of the
emf as measured in procedure (6) compare with these? Why is this?
- What is the distinct advantage of using a potentiometer to measure
the value of an emf of a source over other methods?
- A Potentiometer was used to measure the emf E and internal
resistance r of an unknown battery. With a standard cell of emf
1.0195 V, the balance Position was at 22.1 cm. With the unknown
battery, the balance position was at 65.0 cm. When a 10 W resistor
was connected directly across the battery, the balance position
dropped to 54.2 cm. What were the values of E and r for the battery?
© Dr. John Askill, 1999.
webmaster